After-sales Service: | Yes |
---|---|
Warranty: | 1 Year Warranty |
Display Type: | Digital Only |
FLUKE-1555 10K Insul Tester Kit w/IR3000
Fluke 1555 Insulation Resistance Tester Kit
The extraordinary Fluke 1555 now comes in a kit. With this kit you will not only get this wide-range testing insulation tester, you will also receive ruggedized alligator clips, a Certificate of Calibration, and an IP67 hard case. The Fluke 1555 insulation resistance tester gives you the opportunity of digital insulation testing up to 10kV which makes it optimal for testing a wide range of high-voltage equipment. While the Fluke 1555 can now conduct the entire range of test voltages specified in IEEE 43-2000, it also has a 3-year warranty and CAT IV 600 V safety rating. The 1555 is a perfect tool for preventative or predictive maintenance programs that are designed to identify potential equipment failures before they occur.
Electrical Specifications | The tester's accuracy is specified for one year after calibration at operating temperatures of 0 to 35°C For operating temperatures outside the range (-20 to 0°C and 35 to 50°C), add ±0.25% per °C, except on the 20% bands add ±1% per °C |
Insulation Resistance Measurement | |
250 V | Range: <250 kΩ 250 kΩ to 5 GΩ 5 to 50 GΩ >50 GΩ Accuracy (± reading): Unspecified 5% 20% Unspecified |
500 V | Range: <500 kΩ 500 kΩ to 10 GΩ 10 to 100 GΩ >100 GΩ Accuracy (± reading): Unspecified 5% 20% Unspecified |
1000 V | Range: <1 MΩ 1 MΩ to 20 GΩ 20 to 200 GΩ >200 GΩ Accuracy (± reading): Unspecified 5% 20% Unspecified |
2500 V | Range: <2.5 MΩ 2.5 MΩ to 50 GΩ 50 to 500 GΩ >500 GΩ Accuracy (± reading): Unspecified 5% 20% Unspecified |
5000 V | Range: <5 MΩ 5 MΩ to 100 GΩ 100 GΩ to 1 TΩ >1 TΩ Accuracy (± reading): Unspecified 5% 20% Unspecified |
10,000 V | Range: <10 MΩ 10 MΩ to 200 GΩ 200 GΩ to 2 TΩ >2 TΩ Accuracy (± reading): Unspecified 5% 20% Unspecified |
Bar Graph Range | 0 to 1 TΩ |
Insulation Test Voltage Accuracy | -0%, +10% at 1 mA load current |
Induced AC Mains Current Rejection | 2 mA maximum |
Capacitive Load | Charging rate: 5 seconds per µF Discharge rate: 1.5 seconds per µF |
Leakage Current Measurement | Range: 1 nA to 2 mA Accuracy: ±(5% + 2 nA) |
Capacitance Measurement | Range: 0.01 to 15 µF Accuracy: ±(15% rdg + 0.03 µF) |
Timer | Range: 0 to 99 minutes Resolution setting: 1 minute Resolution indication: 1 second |
Live Circuit Warning | Warning range: 30 to 660 V AC/DC, 50/60 Hz Voltage accuracy: ±(15% + 2 V) |
General Specifications | |
Software | FlukeView Forms™ basic software requires a PC running Windows XP, Vista, 7, or 8 |
Display | 2.9 x 4" (75 x 105 mm) |
Power | 12 V lead-acid rechargeable battery, Yuasa NP2.8-12 Charger input (AC): 85 to 250 V AC 50/60 Hz 20 VA |
Environmental | Operating temperature: -4 to 122°F (-20 to 50°C) Storage temperature: -4 to 149°F (-20 to 65°C) Humidity: 80% to 31°C decreasing linearly to 50% at 50°C Altitude: 6561.68' (2000 m) Enclosure sealing: IP40 |
Input Overload Protection | 1000 V AC |
EMC | EN 61326 |
Safety Compliance | IEC 61010-1: Overvoltage category II, Pollution Degree 2 IEC 61010-2-030: Measurement 600 V CAT IV/1000 V CAT III IEC 61557-1, IEC 61557-2 |
Dimensions | 10.6 x 10.9 x 6.3" (269 x 277 x 160 mm) |
Weight | 8.2 lbs (3.7 kg) |